Product performance complaint received on volatiles
Sep 18, 2025 08:25 PM
Hello Andrew,
Please see my responses below. Thank you for providing the link to QC technical support. I will submit any future concerns through that portal.
Field Required Customer Entry
Customer Identification Number (CID) 021415
Contact Name Katherine Wesson
Contact Email katherine.s.wesson@kp.org
Sales Order #
(or) Lot#
(or) Part#
*Two of the three are required Sales Order # 54934
Lot # D5369
Product # 88119
What happened?
(e.g., Part number 1234 was shipped) Degradation of analyte in quality control material. Although a steady downward trend is observed in the four analytes monitored in this control (methanol, ethanol, acetone, isopropanol), acetone displays a significant degradation after only 4.5 months of use.
What should have happened?
(e.g., part number XYZ should have been shipped) Normally, once established, the QC range for the volatiles control requires no adjustment for the life of the lot.
Additional Comments
(e.g., steps already taken to troubleshoot, additional notes, or preferred resolutions) Three calibrator lots have been in use since the QC material was implemented in production. Only one instrument runs this assay; maintenance is up to date, a PM was done recently on 08/25/25 - FSE did not find anything. CAP survey material was repeated, and the repeat results were comparable to the original results, which indicates to us there is no issue with instrument performance, further supporting our suspicion that the QC material is degrading. Samples of the QC material were sent out to Quest for comparison, and their data also showed values in the range similar to what we are recovering.
Regards,
Katherine Wesson
Toxicology QA/QC Supervisor
Kaiser Regional Laboratory
914 Marina Way S
Richmond, CA 94804
Office Phone: 510-231-5172
Email: katherine.s.wesson@kp.org
________________________________________
From: Andrew Hartmann <ahartmann@utak.com>
Sent: Thursday, September 18, 2025 9:15 AM
To: Katherine S Wesson <Katherine.S.Wesson@kp.org>; Kevin Kopp <kevin@utak.com>
Cc: Christine M Mulakken <Christine.M.Giang@kp.org>; Sarah B Shugarts <Sarah.B.Shugarts@kp.org>; Jay L Salas <Jay.Salas@kp.org>; Julie Cruz <Julie.Cruz@kp.org>; Melinda K. Wade <Melinda.K.Wade@kp.org>
Subject: RE: Volatiles Control - Lot D5369
Caution: This email came from outside Kaiser Permanente. Do not open attachments or click on links if you do not recognize the sender.
________________________________________
Hello Katherine,
Thank you for reaching out to UTAK. We’re here to help and want to resolve this quickly. For the most expedient support (or troubleshooting guides), you may submit your request here: UTAK QC Tech Support.
Could you please share the following details so we can properly log and investigate your concern?
Field Required Customer Entry
Customer Identification Number (CID)
Contact Name
Contact Email
Sales Order #
(or) Lot#
(or) Part#
*Two of the three are required
What happened?
(e.g., Part number 1234 was shipped)
What should have happened?
(e.g., part number XYZ should have been shipped)
Additional Comments
(e.g., steps already taken to troubleshoot, additional notes, or preferred resolutions)
Once we have this information, we can move forward quickly with next steps.
We appreciate your partnership and look forward to resolving this promptly.
Best regards,
Andrew Hartmann
Business Development Manager
–
25020 Avenue Tibbitts
Valencia, CA 91355
–
D +1.661.567.0667
P +1.888.UTAK.LAB
F +1.661.294.9272
–
Book a Meeting with Me
https://calendly.com/ahartmann-utak/30min
utak.com
From: Katherine S Wesson <Katherine.S.Wesson@kp.org>
Sent: Wednesday, September 17, 2025 6:05 PM
To: Support@UTAK <support@utak.com>
Cc: Christine M Mulakken <Christine.M.Giang@kp.org>; Sarah B Shugarts <Sarah.B.Shugarts@kp.org>; Jay L Salas <Jay.Salas@kp.org>; Julie Cruz <Julie.Cruz@kp.org>; Melinda K. Wade <Melinda.K.Wade@kp.org>
Subject: Volatiles Control - Lot D5369
Importance: High
Hello UTAK,
The stability of Volatiles Control Lot D5369 has not been comparable to previous lots. All analytes have been treding down, most especially acetone. This lot has been in use since 04/28/25 and we have observed a steady drift in recovery over time. The verified value for this lot is 75 mg/dL, but our current set mean is 64.5 mg/dL. The calibrator material has been replaced multiple times since this control started. The instrument has been PM'd and verified as acceptable by the vendor's engineer.
We store this material at the recommended refrigerator temperature, stored at 2-8 oC, and the refrigerator is montired 24/7 with no recorded outliers. The control is reconstituted per insert instructions.
Have any customers contacted UTAK about this material? Would it be possible to review the manufacturing records for this control material? Is the storage condition of 2-8oC still correct?
Regards,
Katherine Wesson
Toxicology QA/QC Supervisor