Activity

Email
Additional details on the product performance issue
Sep 30, 2025 01:16 PM
Hi Victor, Thank you for sharing this additional information. @Kevin Kopp from our technical team is reviewing and will reply within the next 1-2 business days with an update. Thank you, Andrew Hartmann Business Development Manager – 25020 Avenue Tibbitts Valencia, CA 91355 – D +1.661.567.0667 P +1.888.UTAK.LAB F +1.661.294.9272 – Book a Meeting with Me https://calendly.com/ahartmann-utak/30min utak.com From: Victor Asirvatham <victor.asirvatham@genesisreferencelabs.com> Sent: Monday, September 29, 2025 5:35 PM To: Andrew Hartmann <ahartmann@utak.com> Cc: Kevin Kopp <kevin@utak.com> Subject: Re: COA - CID#100461 Hi Andrew, Take a look at the LJ chart before and after we lot # D4778 into use on 08/01/25. This is hydrocodone. Several others are also giving around 30% recoveries compared to the nominal values. Field Required Customer Entry Customer Identification Number (CID) 100461 Contact Name Victor Asirvatham Contact Email Victor.asirvatham@genesisreferencelabs.com Sales Order # (or) Lot# (or) Part# *Two of the three are required Lot #D4778 Part # 50701 What happened? (e.g., Part number 1234 was shipped) Recovery doesn’t the IFU sheet What should have happened? (e.g., part number XYZ should have been shipped) Recoveries should be within 20% from the nominal values Additional Comments (e.g., steps already taken to troubleshoot, additional notes, or preferred resolutions) Previous lot DAU high 2 were good
History
  • Created Sep 30, 2025 01:17 PM
    Activity created