Serum calibrator quote requested
Oct 03, 2025 08:08 PM
Thanks Matt!
I’ll circle back if the team has any follow up questions, otherwise I’ll connect with you when the quotes are ready to review. Should have them early next week.
Thank you,
Andrew Hartmann
Business Development Manager
–
25020 Avenue Tibbitts
Valencia, CA 91355
–
D +1.661.567.0667
P +1.888.UTAK.LAB
F +1.661.294.9272
–
Book a Meeting with Me
https://calendly.com/ahartmann-utak/30min
utak.com
From: Matthew Bown <MaBown@pathgroup.com>
Sent: Friday, October 3, 2025 1:13 PM
To: Andrew Hartmann <ahartmann@utak.com>; Jordan B. Wesler <JBWesler@pathgroup.com>
Subject: Re: [EXTERNAL] RE: QC separation
Hi Andrew,
No problem. Please see below:
analyte cutoff level 1 level 2 level 3 level 4 level 5 level 6 level 7
6-Acetylmorphine 2 1 1.5 2 5 10 20 40
7-amino-clonazepam 5 2.5 3.75 5 12.5 25 50 100
Alprazolam 5 2.5 3.75 5 12.5 25 50 100
Amphetamine 20 10 15 20 50 100 200 400
Benzoylecgonine 20 10 15 20 50 100 200 400
Buprenorphine 1 0.5 0.75 1 2.5 5 10 20
Chlordiazepoxide 20 10 15 20 50 100 200 400
Clonazepam 5 2.5 3.75 5 12.5 25 50 100
Codeine 2 1 1.5 2 5 10 20 40
Diazepam 5 2.5 3.75 5 12.5 25 50 100
EDDP 10 5 7.5 10 25 50 100 200
Midazolam 20 10 15 20 50 100 200 400
alpha-hydroxymidazolam 20 10 15 20 50 100 200 400
Hydrocodone-s 2 1 1.5 2 5 10 20 40
Hydromorphone 2 1 1.5 2 5 10 20 40
Hydroxyalprazolam 5 2.5 3.75 5 12.5 25 50 100
Lorazepam 20 10 15 20 50 100 200 400
11-nor-9-Carboxy-Δ9-THC 5 2.5 3.75 5 12.5 25 50 100
MDA 20 10 15 20 50 100 200 400
MDEA 20 10 15 20 50 100 200 400
MDMA 20 10 15 20 50 100 200 400
Methadone 2 1 1.5 2 5 10 20 40
Methamphetamine 20 10 15 20 50 100 200 400
Nicotine 5 2.5 3.75 5 12.5 25 50 100
Morphine 2 1 1.5 2 5 10 20 40
Norbuprenorphine 1 0.5 0.75 1 2.5 5 10 20
Nordiazepam 20 10 15 20 50 100 200 400
Cotinine 5 2.5 3.75 5 12.5 25 50 100
Oxazepam 20 10 15 20 50 100 200 400
Oxycodone 2 1 1.5 2 5 10 20 40
Oxymorphone 2 1 1.5 2 5 10 20 40
Phencyclidine 10 5 7.5 10 25 50 100 200
Temazepam 20 10 15 20 50 100 200 400
Butalbital 50 25 37.5 50 125 250 500 1000
Pentobarbital 50 25 37.5 50 125 250 500 1000
Phenobarbital 50 25 37.5 50 125 250 500 1000
Best,
Matt Bown, MS
LCMS Application Specialist
1010 Airpark Center Drive
Nashville, TN 37217
Mobile: 615-393-2316
Book time to meet with me
________________________________________
From: Andrew Hartmann <ahartmann@utak.com>
Sent: Friday, October 3, 2025 12:00 PM
To: Matthew Bown <MaBown@pathgroup.com>; Jordan B. Wesler <JBWesler@pathgroup.com>
Subject: RE: [EXTERNAL] RE: QC separation
Hi Matt,
I am having trouble finding this analyte list. Could you send it over along with what the cut-off concentration is for each. With that we can figure out all the levels.
What you laid out as far as design follows what we typically see so from a design standpoint I think you are pretty spot on.
Thank you,
Andrew Hartmann
Business Development Manager
–
25020 Avenue Tibbitts
Valencia, CA 91355
–
D +1.661.567.0667
P +1.888.UTAK.LAB
F +1.661.294.9272
–
Book a Meeting with Me
https://calendly.com/ahartmann-utak/30min
utak.com
From: Matthew Bown <MaBown@pathgroup.com>
Sent: Friday, October 3, 2025 11:16 AM
To: Andrew Hartmann <ahartmann@utak.com>; Jordan B. Wesler <JBWesler@pathgroup.com>
Subject: Re: [EXTERNAL] RE: QC separation
Hey Andrew,
Apologies for the delay getting back to you.
We'll require separate lots used for Cals/QC.
For the single calibrator at 20x cutoffs:
1. What matrix do you want this in
a. Serum or plasma (whichever is cheaper)
2. What fill size per vial do you want?
a. 1000 µL
3. How many vials would you use in a year, ball park estimate
a. 104.
4. We need a list of analytes and target concentration
a. Are you able to pull my latest order of serum drugs? Its the same analytes/cutoffs as that.
For the a 7 point calibration curve:
1. What matrix do you want this in
a. Serum or plasma (whichever is cheaper)
2. What fill size per vial do you want?
a. 300 µL.
3. How many vials would you use in a year, ball park estimate
a. 104 per calibrator.
4. We need a list of analytes and target concentration
a. Are you able to pull my latest order of serum drugs? Its the same analytes/cutoffs.
b. The cal levels (7) will go 0.5x CO, 0.75x CO, CO, 2.5x CO, 5x CO, 10x CO, 20x CO.
What is a common design you see customers order for toxicology applications like this? Can you include that design with our drugs/cutoffs? I may not be going about this the most efficient way..